Samenvatting
PURPOSE: To report intra- and interinstrument calibration methods for corneal backscatter analysis by in vivo confocal microscopy.
METHODS: Applicability of two reference standards was evaluated for corneal backscatter calibration. Repeated measurements of four concentrations of AMCO Clear (GFS Chemicals, Inc., Powell, OH) suspension and three transparencies (26%, 49%, and 65%) of polymethylmethacrylate (PMMA) slabs were performed to assess image intensity acquisition in a wide backscatter range. Intra- and intersession repeatability and lot-to-lot variation were determined for both standards. The effect of light intensity (LI) variation on image intensity acquisition was evaluated by examination of PMMA slabs with nonreference (60% and 80%) and reference (72%) LIs. Both reference standards were implemented in the protocol. Intrainstrument calibration was verified by measuring three normal corneas with 60%, 72%, and 80% LIs. Interinstrument calibration was tested by measuring PMMA slabs on a second, similar confocal microscope.
RESULTS: AMCO Clear was used to express image intensity in absolute scatter units (SU), whereas the 49% transparent PMMA slab showed best repeatability, without image saturation, to adjust for LI variation. Intrainstrument calibration for LI variation reduced mean differences from -38.3% to 1.7% (60% LI) and from 33.9% to -0.6% (80% LI). The mean difference between similar microscopes decreased from 18.4% to 1.2%, after calibration of the second microscope.
CONCLUSIONS: Large interinstrument differences necessitate calibration of corneal backscatter measurements. With AMCO Clear suspension and PMMA slabs, standardization was achieved in a wide backscatter range corresponding to normal and opaque corneas. These methods can easily be applied in ophthalmic practice.
| Originele taal-2 | Engels |
|---|---|
| Pagina's (van-tot) | 2136-46 |
| Aantal pagina's | 11 |
| Tijdschrift | Investigative ophthalmology & visual science |
| Volume | 52 |
| Nummer van het tijdschrift | 5 |
| DOI's | |
| Status | Gepubliceerd - 5 apr 2011 |
Vingerafdruk
Duik in de onderzoeksthema's van 'Wide-range calibration of corneal backscatter analysis by in vivo confocal microscopy'. Samen vormen ze een unieke vingerafdruk.Citeer dit
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver