TY - JOUR
T1 - Wide-range calibration of corneal backscatter analysis by in vivo confocal microscopy
AU - Hillenaar, Toine
AU - Sicam, Victor Arni D P
AU - Vermeer, Koenraad A
AU - Braaf, Boy
AU - Remeijer, Lies
AU - Cals, Roger H H
AU - de Boer, Johannes F
PY - 2011/4/5
Y1 - 2011/4/5
N2 - PURPOSE: To report intra- and interinstrument calibration methods for corneal backscatter analysis by in vivo confocal microscopy.METHODS: Applicability of two reference standards was evaluated for corneal backscatter calibration. Repeated measurements of four concentrations of AMCO Clear (GFS Chemicals, Inc., Powell, OH) suspension and three transparencies (26%, 49%, and 65%) of polymethylmethacrylate (PMMA) slabs were performed to assess image intensity acquisition in a wide backscatter range. Intra- and intersession repeatability and lot-to-lot variation were determined for both standards. The effect of light intensity (LI) variation on image intensity acquisition was evaluated by examination of PMMA slabs with nonreference (60% and 80%) and reference (72%) LIs. Both reference standards were implemented in the protocol. Intrainstrument calibration was verified by measuring three normal corneas with 60%, 72%, and 80% LIs. Interinstrument calibration was tested by measuring PMMA slabs on a second, similar confocal microscope.RESULTS: AMCO Clear was used to express image intensity in absolute scatter units (SU), whereas the 49% transparent PMMA slab showed best repeatability, without image saturation, to adjust for LI variation. Intrainstrument calibration for LI variation reduced mean differences from -38.3% to 1.7% (60% LI) and from 33.9% to -0.6% (80% LI). The mean difference between similar microscopes decreased from 18.4% to 1.2%, after calibration of the second microscope.CONCLUSIONS: Large interinstrument differences necessitate calibration of corneal backscatter measurements. With AMCO Clear suspension and PMMA slabs, standardization was achieved in a wide backscatter range corresponding to normal and opaque corneas. These methods can easily be applied in ophthalmic practice.
AB - PURPOSE: To report intra- and interinstrument calibration methods for corneal backscatter analysis by in vivo confocal microscopy.METHODS: Applicability of two reference standards was evaluated for corneal backscatter calibration. Repeated measurements of four concentrations of AMCO Clear (GFS Chemicals, Inc., Powell, OH) suspension and three transparencies (26%, 49%, and 65%) of polymethylmethacrylate (PMMA) slabs were performed to assess image intensity acquisition in a wide backscatter range. Intra- and intersession repeatability and lot-to-lot variation were determined for both standards. The effect of light intensity (LI) variation on image intensity acquisition was evaluated by examination of PMMA slabs with nonreference (60% and 80%) and reference (72%) LIs. Both reference standards were implemented in the protocol. Intrainstrument calibration was verified by measuring three normal corneas with 60%, 72%, and 80% LIs. Interinstrument calibration was tested by measuring PMMA slabs on a second, similar confocal microscope.RESULTS: AMCO Clear was used to express image intensity in absolute scatter units (SU), whereas the 49% transparent PMMA slab showed best repeatability, without image saturation, to adjust for LI variation. Intrainstrument calibration for LI variation reduced mean differences from -38.3% to 1.7% (60% LI) and from 33.9% to -0.6% (80% LI). The mean difference between similar microscopes decreased from 18.4% to 1.2%, after calibration of the second microscope.CONCLUSIONS: Large interinstrument differences necessitate calibration of corneal backscatter measurements. With AMCO Clear suspension and PMMA slabs, standardization was achieved in a wide backscatter range corresponding to normal and opaque corneas. These methods can easily be applied in ophthalmic practice.
KW - Calibration
KW - Cornea/chemistry
KW - Humans
KW - Light
KW - Microscopy, Confocal/instrumentation
KW - Polymethyl Methacrylate/chemistry
KW - Reference Standards
KW - Scattering, Radiation
KW - Styrene/chemistry
KW - Vinyl Compounds/chemistry
U2 - 10.1167/iovs.10-6314
DO - 10.1167/iovs.10-6314
M3 - Article
C2 - 21071733
SN - 0146-0404
VL - 52
SP - 2136
EP - 2146
JO - Investigative ophthalmology & visual science
JF - Investigative ophthalmology & visual science
IS - 5