Parallel line scanning ophthalmoscope for retinal imaging

  • Kari V Vienola
  • , Mathi Damodaran
  • , Boy Braaf
  • , Koenraad A Vermeer
  • , Johannes F de Boer

Onderzoeksoutput: Bijdrage aan tijdschrift › Artikel › Onderzoek › peer review

Samenvatting

A parallel line scanning ophthalmoscope (PLSO) is presented using a digital micromirror device (DMD) for parallel confocal line imaging of the retina. The posterior part of the eye is illuminated using up to seven parallel lines, which were projected at 100 Hz. The DMD offers a high degree of parallelism in illuminating the retina compared to traditional scanning laser ophthalmoscope systems utilizing scanning mirrors. The system operated at the shot-noise limit with a signal-to-noise ratio of 28 for an optical power measured at the cornea of 100 μW. To demonstrate the imaging capabilities of the system, the macula and the optic nerve head of a healthy volunteer were imaged. Confocal images show good contrast and lateral resolution with a 10°×10° field of view.

Originele taal-2Engels
Pagina's (van-tot)5335-8
Aantal pagina's4
TijdschriftOptics Letters
Volume40
Nummer van het tijdschrift22
DOI's
StatusGepubliceerd - 15 nov 2015