TY - JOUR
T1 - Effects of inadequate anterior segment compensation on measurements with scanning laser polarimetry
AU - Reus, Nicolaas J
AU - van Koolwijk, Leonieke M E
AU - Lemij, Hans G
PY - 2006/2/14
Y1 - 2006/2/14
N2 - The effects of poor anterior segment compensation on scanning laser polarimetry measurements of the retinal nerve fiber layer (RNFL) were systematically explored. A prototype scanning laser polarimeter with an adjustable compensator to neutralize anterior segment birefringence was used. By systematically varying the magnitude and axis of anterior segment compensation in a healthy and a glaucomatous eye, marked changes were observed in RNFL appearance: the healthy eye could appear to have glaucomatous damage, whereas the glaucomatous eye could appear to have a thicker and healthier RNFL. Even small amounts of uncompensated corneal birefringence, which may occur in routine clinical use, resulted in apparent changes in RNFL morphology. Knowledge of this effect is important for clinicians when using scanning laser polarimetry in clinical practice.
AB - The effects of poor anterior segment compensation on scanning laser polarimetry measurements of the retinal nerve fiber layer (RNFL) were systematically explored. A prototype scanning laser polarimeter with an adjustable compensator to neutralize anterior segment birefringence was used. By systematically varying the magnitude and axis of anterior segment compensation in a healthy and a glaucomatous eye, marked changes were observed in RNFL appearance: the healthy eye could appear to have glaucomatous damage, whereas the glaucomatous eye could appear to have a thicker and healthier RNFL. Even small amounts of uncompensated corneal birefringence, which may occur in routine clinical use, resulted in apparent changes in RNFL morphology. Knowledge of this effect is important for clinicians when using scanning laser polarimetry in clinical practice.
KW - Anterior Eye Segment/anatomy & histology
KW - Birefringence
KW - Diagnostic Techniques, Ophthalmological/instrumentation
KW - Humans
KW - Lasers
KW - Optic Nerve/anatomy & histology
KW - Reference Values
KW - Reproducibility of Results
KW - Retinal Ganglion Cells/cytology
M3 - Article
C2 - 16468552
SN - 1542-8877
VL - 37
SP - 54
EP - 57
JO - Ophthalmic surgery, lasers & imaging : the official journal of the International Society for Imaging in the Eye
JF - Ophthalmic surgery, lasers & imaging : the official journal of the International Society for Imaging in the Eye
IS - 1
ER -