T1 - A Randomized Trial of a Schlemm's Canal Microstent with Phacoemulsification for Reducing Intraocular Pressure in Open-Angle Glaucoma
AU - Pfeiffer, Norbert
AU - Garcia-Feijoo, Julian
AU - Martinez-de-la-Casa, Jose M
AU - Larrosa, Jose M
AU - Fea, Antonio
AU - Lemij, Hans
AU - Gandolfi, Stefano
AU - Schwenn, Oliver
AU - Lorenz, Katrin
AU - Samuelson, Thomas W
N1 - Copyright © 2015 American Academy of Ophthalmology. Published by Elsevier Inc. All rights reserved.
PY - 2015/7
Y1 - 2015/7
N2 - PURPOSE: To assess the safety and effectiveness of the Hydrus Microstent (Ivantis, Inc, Irvine, CA) with concurrent cataract surgery (CS) for reducing intraocular pressure (IOP) in open-angle glaucoma (OAG).DESIGN: Prospective, multicenter, randomized, single-masked, controlled clinical trial.PARTICIPANTS: One hundred eyes from 100 patients 21 to 80 years of age with OAG and cataract with IOP of 24 mmHg or less with 4 or fewer hypotensive medications and a washed-out diurnal IOP (DIOP) of 21 to 36 mmHg.METHODS: On the day of surgery, patients were randomized 1:1 to undergo CS with the microstent or CS alone. Postoperative follow-up was at 1 day, 1 week, and 1, 3, 6, 12, 18, and 24 months. Washout of hypotensive medications was repeated at 12 and 24 months.MAIN OUTCOME MEASURES: Response to treatment was defined as a 20% or more decrease in washed out DIOP at 12 and 24 months of follow-up compared with baseline. Mean DIOP at 12 and 24 months, the proportion of subjects requiring medications at follow-up, and the mean number of medications were analyzed. Safety measures included change in visual acuity, slit-lamp observations, and adverse events.RESULTS: The proportion of patients with a 20% reduction in washed out DIOP was significantly higher in the Hydrus plus CS group at 24 months compared with the CS group (80% vs. 46%; P = 0.0008). Washed out mean DIOP in the Hydrus plus CS group was significantly lower at 24 months compared with the CS group (16.9±3.3 mmHg vs. 19.2±4.7 mmHg; P = 0.0093), and the proportion of patients using no hypotensive medications was significantly higher at 24 months in the Hydrus plus CS group (73% vs. 38%; P = 0.0008). There were no differences in follow-up visual acuity between groups. The only notable device-related adverse event was focal peripheral anterior synechiae (1-2 mm in length). Otherwise, adverse event frequency was similar in the 2 groups.CONCLUSIONS: Intraocular pressure was clinically and statistically significantly lower at 2 years in the Hydrus plus CS group compared with the CS alone group, with no differences in safety.
AB - PURPOSE: To assess the safety and effectiveness of the Hydrus Microstent (Ivantis, Inc, Irvine, CA) with concurrent cataract surgery (CS) for reducing intraocular pressure (IOP) in open-angle glaucoma (OAG).DESIGN: Prospective, multicenter, randomized, single-masked, controlled clinical trial.PARTICIPANTS: One hundred eyes from 100 patients 21 to 80 years of age with OAG and cataract with IOP of 24 mmHg or less with 4 or fewer hypotensive medications and a washed-out diurnal IOP (DIOP) of 21 to 36 mmHg.METHODS: On the day of surgery, patients were randomized 1:1 to undergo CS with the microstent or CS alone. Postoperative follow-up was at 1 day, 1 week, and 1, 3, 6, 12, 18, and 24 months. Washout of hypotensive medications was repeated at 12 and 24 months.MAIN OUTCOME MEASURES: Response to treatment was defined as a 20% or more decrease in washed out DIOP at 12 and 24 months of follow-up compared with baseline. Mean DIOP at 12 and 24 months, the proportion of subjects requiring medications at follow-up, and the mean number of medications were analyzed. Safety measures included change in visual acuity, slit-lamp observations, and adverse events.RESULTS: The proportion of patients with a 20% reduction in washed out DIOP was significantly higher in the Hydrus plus CS group at 24 months compared with the CS group (80% vs. 46%; P = 0.0008). Washed out mean DIOP in the Hydrus plus CS group was significantly lower at 24 months compared with the CS group (16.9±3.3 mmHg vs. 19.2±4.7 mmHg; P = 0.0093), and the proportion of patients using no hypotensive medications was significantly higher at 24 months in the Hydrus plus CS group (73% vs. 38%; P = 0.0008). There were no differences in follow-up visual acuity between groups. The only notable device-related adverse event was focal peripheral anterior synechiae (1-2 mm in length). Otherwise, adverse event frequency was similar in the 2 groups.CONCLUSIONS: Intraocular pressure was clinically and statistically significantly lower at 2 years in the Hydrus plus CS group compared with the CS alone group, with no differences in safety.
KW - Adult
KW - Aged
KW - Aged, 80 and over
KW - Antihypertensive Agents/administration & dosage
KW - Female
KW - Follow-Up Studies
KW - Glaucoma Drainage Implants
KW - Glaucoma, Open-Angle/physiopathology
KW - Humans
KW - Intraocular Pressure/physiology
KW - Lens Implantation, Intraocular
KW - Limbus Corneae/surgery
KW - Male
KW - Middle Aged
KW - Phacoemulsification
KW - Prospective Studies
KW - Prosthesis Implantation
KW - Single-Blind Method
KW - Stents
KW - Tonometry, Ocular
KW - Visual Acuity/physiology
KW - Visual Fields
KW - Young Adult
U2 - 10.1016/j.ophtha.2015.03.031
DO - 10.1016/j.ophtha.2015.03.031
M3 - Article
C2 - 25972254
SN - 0161-6420
VL - 122
SP - 1283
EP - 1293
JO - Ophthalmology
JF - Ophthalmology
IS - 7
ER -